beautypg.com

HEIDENHAIN ND 710 v.3 User Manual

Page 37

background image

38

Operating

Parameters

P41.1 to P41.3

Linear axis error compensation

Input range (µm):

-99999 to +99999

Factory default setting:

0

Example: Displayed length

Ld = 620.000 mm

Actual length (as determined for example with

the VM 101 from HEIDENHAIN)

La = 619.876 mm

Difference

DL = L

a

– L

d

= – 124 µm

Compensation factor k:

k = DL/Ld = – 124 µm/0.62 m = – 200 [µm/m]

P42.1 to P42.3 Backlash compensation
Input range (mm):

+9.999 to –9.999

Default setting:

0.000 = no backlash
compensation

A change in direction might cause a clearance between
rotary encoder and table. This is referred to as backlash.
Positive backlash: The rotary encoder is ahead of the table.
The distance traversed by the table is too short (entry of
positive value).
Negative backlash: The rotary encoder follows the table. The
distance traversed by the table is too large (entry of negative
value).

P43.1 to P43.3 Reference marks
One reference mark

SINGLE REF.M.

Distance-coded with 500 • SP

500

SP

Distance-coded with 1000 • SP

1000

SP

Distance-coded with 2000 • SP

2000

SP

Distance-coded with 5000 • SP

5000

SP

(SP: signal period)

P44.1 to P44.3 Reference mark evaluation
Evaluation

REF. X ON

No evaluation

REF. X OFF

P45.1 to P45.3 Encoder monitoring
Amplitude and frequency

monitoring

ALARM ON

No monitoring

ALARM OFF

P48.1 to P48.3 Activate axis display
Axis display active

AXIS ON

Not active

AXIS OFF

P70 Number of datums
2 datums

2 DATUM PT.

9 datums

9 DATUM PT.

P80 Function of the CL key
Reset to zero with CL

CL...RESET

No reset to zero with CL

CL......OFF

P81 Function of the R+/- key
The R+/- key divides the actual

R+/- 1/2

value by two
No division of the actual value

R+/- OFF

Bateil2.pm6

27.03.2002, 11:33

38

This manual is related to the following products: