beautypg.com

Error messages – HEIDENHAIN ND 710 v.3 User Manual

Page 29

background image

29

To erase error messages:

After you have removed the cause of error:

➤ Press the CL key.

Error Messages

Message

Cause and effect

SIGNAL X

Encoder signal is too small,

e.g. when an encoder is

contaminated.

PROB. ERROR

Before touching off on the

workpiece, the tool must move

by a distance of at least 0.2 mm.

ERR. REF. X

The spacing of the reference

marks as defined in P43 is not

the same as the actual spacing.

FRQ. ERR. X

The input frequency for this

encoder input is too high. This

can occur when the scale is

moved too fast.

ERR. MEMORY

Check sum error: Check the

datum, operating parameters and

compensation values for multi-

point axis error compensation.

If the error recurs, contact your

service agency!