List of operating parameters
P00 CODE Enter the code number:
9 51 48:
Change protected operating parameters
66 55 44:
Display the software version (X display)
Display the date of release (Y display)
10 52 96:
Multipoint axis error compensation
P01 Unit of measure
1)
Display in millimeters
MM
Display in inches
INCH
P03.1 to P03.3 Radius/diameter display
1)
Display position value as radius
RADIUS
Display position value as diameter
DIAMETER
P11 Activate scaling factor
1)
Active
SCALING ON
Not active
SCALING OFF
P12.1 to P12.3 Define scaling factor
1)
Enter a scaling factor separately for each axis:
Entry value > 1: workpiece will grow
Entry value = 1: workpiece will remain the same size
Entry value < 1: workpiece will shrink
Input range:
0.100000 to 9.999999
Factory default setting:
1.000000
P30.1 to P30.3 Counting direction
Positive counting direction with
positive direction of traverse
DIRECT. POS
Negative counting direction with
positive direction of traverse
DIRECT. NEG
P31.1 to P31.3 Signal period of the encoder
Input range:
0.00000001 to 99999.9999 µm
Default setting:
20 µm
P33.1 to P33.3 Counting mode
0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9
0 - 2 - 4 - 6 - 8
0 - 5
P38.1 to P38.3 Decimal places
1 / 2 / 3 / 4 (up to 6 with inch display)
P40.1 to P40.3 Select type of axis error compensation
No axis error compensation
COMP. OFF
Linear error compensation active,
multipoint error comp. not active
COMP. LIN
Multipoint error compensation active,
linear error compensation not active COMP. MULTI