Operating

Parameters

Operating Parameters

Operating parameters allow you to modify the operating

characteristics of your display unit and define the evaluation

of the encoder signals. Operating parameters that can be

changed by the user are called user parameters, and can be

accessed with the SPEC FCT key and the dialog

“PARAMETER” (user parameters are identified as such in the

parameter list). The full range of parameters can only be

accessed through the dialog “CODE“ and by entering 95148.

Operating parameters are designated by the letter P and a

number. Example: P11. The parameter designation is shown

in the input field when you select it with the DATUM and

ENT key in the X display. The parameter setting is shown in

the Y display.

Some operating parameters have separate values for each

axis. Such parameters have an additional index number from

1 to 3 with the ND 750 and from 1 to 2 with the ND 710.

Example: P12.1 scaling factor, X axis

P12.2 scaling factor, Y axis

P12.3 scaling factor, Z axis (ND 750 only)

The operating parameters are preset before the unit leaves

the factory. These factory settings are indicated in the

parameter list in boldface type.

Entering and changing operating parameters

To access the operating parameters
➤ Press the SPEC FCT key.

➤ Press the SPEC FCT key or 1 2 , until

“PARAMETER” appears in the X display.

➤ Confirm your selection by pressing “ENT.”

To select protected operating parameters
➤ Press the 1 2 key to select user parameter

P00 CODE.

➤ Enter the code number 95148.

➤ Confirm with “ENT.”

To page through the operating parameters
➤ Page forwards by pressing the ENT key.

➤ Page backwards by pressing the 1 2 key.

To change parameter settings
➤ Press the minus key or enter the value and confirm

with the ENT key.

To correct an entry
➤ Press CL: the old value reappears in the input line and

becomes effective again.

To leave the operating parameters
➤ Press the SPEC FCT or CL key.