5 specifications, 5specifications – HEIDENHAIN ND 200 User Manual
Page 22

Installation Instructions
en
ND 200
24
5
Specifications
Device
Housing
Benchtop design, cast-metal housing
Axes
ND 280: 1 axis
ND 287: 2 axes (2nd axis optional)
Display
Visual display
unit
Dialogs and inputs, graphic functions,
graphic positioning aid
Only ND 287: Additional color LCD dis-
play for position values
Status display: Operating mode, axis
(
ND 287 with axis coupling), scaling fac-
tor, compensation, stopwatch, unit of
measure, datum number, soft-key level
Position display and measured-value
display with selectable display resoluti-
on
Display step
Linear axes: 0.5 mm to 0.001 µm, de-
pending on the signal period
Rotary axes: 0.5° to 0.000001°
(00°00’00.1”), depending on the signal
period
Electrical data
Power supply
100 V AC to 240 V AC; 50 Hz to 60 Hz;
max. 30 VA
Line fuse
2 x T500 mA
Encoder interface 15-pin D-sub socket
Incremental HEIDENHAIN encoders
with sinusoidal signals:
11 µA
PP
Input frequency 100 kHz max.,
Cable length 30 m max.
1 V
PP
Input frequency 500 kHz max.,
Cable length 60 m max.
Absolute HEIDENHAIN encoders with
EnDat interface (purely serial)
Cable length 100 m max.
Possible signal periods for linear and ro-
tary encoders:
For rotary encoders:
1 to 999 999.999
For linear encoders:
0.000 000 01 µm to 99 999.9999 µm
Data
interface
RS-232-C/V.24
9-pin D-sub socket
USB data in-
terface
USB Type B (UART) socket
Cable length: max. 5 m
Electrical data
Only ND 287:
Switching
inputs and
outputs
25-pin D-sub socket
Signal level at the inputs:
High: +3.9 V
≤ U ≤ 15 V
Low: –0.5 V
≤ U ≤ 0.9 V; I ≤ 6 mA
Signal level at the outputs:
High: U
≤ +32 V; I ≤ 10 μA
Low: U
≤ +0.4 V; I ≤ 100 mA
Ambient conditions
Operating
temperature
0 °C to 50 °C (32 °F to 122 °F)
Storage
temperature
-40 °C to 85 °C (-40 °F to 185 °F)
Relative air hu-
midity
Annual mean: < 75 %
In exceptional cases: < 90 %
General information
Directives
EMC Directive 2004/108/EC
Low Voltage Directive 2006/95/EC
Pollution de-
gree
II
Protection
EN 60529
Rear panel: IP 40
Front panel: IP 54
Weight
Approx. 2.5 kg
Dimensions
See dimension drawings in the Appendix.
All dimensions are in millimeters [mm].