Installation Instructions

en

ND 200

20

4

Installation

NOTICE

Disturbances caused by lack of grounding or improper

grounding!

Never operate the unit without proper ground
connection.
Connect the ground connector on the rear panel to
the central grounding point of the unit.
Minimum cross section of the conductor: 6 mm

2

.

NOTICE

Risk of damage to internal components!

Do not engage or disengage any connecting elements
while the unit is under power.

Depending on the unit's equipment, the

installation may differ from the procedure

described in this chapter. If the Addendum

shipped with the product contains information

on the installation, the information given in

the Addendum takes precedence over the

information in this chapter.

The safety of any system incorporating the

use of this product is the responsibility of the

assembler or installer of the system.

The illustrations of pin layouts always show the

pin assignments of the connections on the unit,

and not of the connectors.

Personnel requirement

The following steps are only to be performed by

qualified personnel!
For more information: see "Personnel

qualification", page 16.

4.1

Product overview

Rear panel

See Figure "A" on the front fold-out page. The figure

shows the rear of the ND 287. The ND 287 and ND 280 are

identical in their design, except for the additional connection

possibilities provided by the ND 287, which are listed below.

1

Power switch

2

Power connector

3

Line fuse

4

Ground

X1

15-pin D-sub connection for HEIDENHAIN
encoder with

11 µA

PP

interface or

1 V

PP

interface or

EnDat interface (purely serial)

Only ND 287 (option):
Replacing the encoder module by an analog
module for connection of an analog sensor

X31

Serial 9-pin D-sub connection RS-232-C/V.24
for data transmission

X32

Serial USB connection type B (UART) for da-
ta transmission

Only ND 287:
X41

25-pin D-sub connection for switching in-
puts and outputs

X2

Option: Addition of an expansion module
(as for connection X1)

X26(X27) Option: Addition of an Ethernet expansion

module (100baseT) for network connection
via TCP/IP protocol