beautypg.com

Service, Analog output calibration, Service -9 – Emerson Guided Wave Radar Level and Interface Transmitters 3300 User Manual

Page 89: Analog output calibration -9, Rosemount 3300 series, Reference manual, Service analog output calibration

background image

Reference Manual

00809-0100-4811, Rev CA
February 2006

6-9

Rosemount 3300 Series

SERVICE

Analog Output
Calibration

To calibrate the Analog Output current do the following:

1. Start RCT and make sure that the transmitter communicates with the PC

(see Section 4: Installing the RCT software).

2. Open the Advanced section in the RCT workspace Project Bar and click

the Device Commands icon,
or
choose the Device Commands option from the View menu.

3. Open the folder named Diag and double-click the Fixed Current Mode

option.

4. Set the output current to 4 mA.

5. Measure the output current.

6. Open the folder named Details.

7. Choose the Trim DAC Zero option and enter the measured output

current.

8. In the Diag folder double-click the Fixed Current Mode option and set the

output current to 20 mA.

9. Measure the output current.

10. In the Details folder double-click the Trim DAC Gain option and enter the

measured output current.

11. In the Diag folder double-click the Fixed Current Mode option and set the

output current to 0 mA in order to leave the Fixed Current mode.

Fixed Current
Mode

Advanced

W

O

RKS

P

ACE_A

D

VANCED_

F

IXE

DC

URRENT