beautypg.com

Measurement Computing DaqBook 2000 Series User Manual

Page 81

background image

Specifications are subject to change without notice.

DaqBook/2000 and DaqOEM/2000 Series

898695

Specifications 11-3

Thermocouple Input Specifications for DaqBook/2020 Mini-TC Connectors

TC/mV Connectors: 14 Mini-TC connectors, oriented in 2 rows of 7

Inputs: 14 differential TC/mV inputs, open TC detection per channel

TC Types: J, K, T, E, S, R, B, N28, N14; x100 voltage

Input Impedance: 40 M

Ω (differential); 20 MΩ (single-ended)

Input Voltage Range: ± 100 mV

Input Bandwidth: 4 kHz

Input Bias Current: 10 nA (typ)

CMRR: -100 db (typ)

Maximum Working Voltage (Signal + Common Mode): ±10V

Over-Voltage Protection: ±40V

Voltage Accuracy: ±(0.2% of the reading + 50 µV)

Temperature Coefficient for Voltage Accuracy: 10ppm for every ˚C outside the range of 0° to 50°C

TC Accuracy: Valid for one year, 18° to 28°C, see the following table.

Temperature Coefficient for Type T: ±0.03°C per °C ambient outside the range of 18 to 28˚C

Minimum Resolution: 0.1˚C for all TC types

TC Accuracy

at Measurement Temperature in °C (±°C)

Type

Min

Max

-100

0

100

300

500

700

900

1100

1400

J

-200 760 0.8

0.7

0.7

0.8

0.9

0.9

— — —

K

-200 1200 0.9

0.8

0.8

0.9

1.1

1.1

1.2

1.3

T

-200 400 0.9

0.8

0.8

0.8

— — — — —

E

-270 650 0.8

0.7

0.7

0.7

0.8

— — — —

S

-50 1768 —

3.1

2.4

2.0

2.0

1.9

2.0

2.1

2.1

R

-50 1768 —

3.1

2.1

2.0

1.9

1.9

1.7

1.9

2.0

B

50

1780

— — — 4.9

3.2

2.8

2.4

2.3

2.0

N28

-270 400 1.2

0.9

0.9

0.9

— — — — —

N14

0 1300 — 0.9

0.9

0.9

1.1

1.1

1.2

1.3

Accuracy conditions:

Exclusive of thermocouple errors

Exclusive of noise

VCM = 0