2-2 Device Overviews

898095

DaqBook/2000 Series and DaqOEM/2000 Series

Overview

DaqBook/2020, /2001, /2005, DaqOEM/2001, and DaqOEM/2005 are high-speed, multi-function, data

acquisition devices for use with Ethernet PCs and Ethernet networks. Each unit connects directly to a PC’s

Ethernet port or network hub, and each features a 16-bit, 200-kHz A/D converter and digital calibration.
DaqOEM/2001 and DaqOEM/2005 devices have no chassis, but are otherwise identical to the

DaqBook/2001 and DaqBook/2005, respectively.
Note that block diagrams are presented in chapter 3 and connectors and pinouts are discussed chapter 4.

Product Features

I/O Connections


For DaqBook/2020
, all input and output signals are accessed via: 37-pin DSUB connectors [P1 and P2],

BNC Connectors, and mini-TC connectors. Chapter 4 includes connector descriptions and pinouts.

For DaqBook/2001, /2005 and DaqOEM/2001, and DaqOEM/2005 all input and output signals are

accessed via the three 37-pin DSUB connectors [P1, P2, and P3]. Chapter 4 includes connector

descriptions and pinouts.

General Features

TTL-level digital I/O lines. For DaqBook/2020 there are 24 TTL-level digital I/O lines

divided into three 8-bit ports on P2. For the other /2000 Series devices there are 40 digital

I/O lines. They are divided into three 8-bit ports on P2 and one 16-bit port on P3.

Four 16-bit counters. Each can accept frequency inputs up to 10 MHz. The counters can be

cascaded into two 32-bit counters.

Two 16-bit timer outputs. Each can generate square waves from 16 Hz to 1 MHz.

Configuration through software only. Device settings are made via software. There are no

hardware settings to be made, for example, there are no device DIP switches to be set or

jumpers to be positioned. However, note that DBK expansion options may require s small

degree of manual set up. Refer to the DBK Option Cards and Modules User’s Manual

(457-0905) for details. A PDF version of the document is included on the installation CD.

Scan Sequencing. Each DaqBook/2020, /2001, /2005, DaqOEM/2001, and DaqOEM/2005

has a scan sequencer lets you select up to 16384 channel/range combinations. Scan rates are

programmable on a per channel basis to 5 µs/channel, 10 µs/channel, or 1000 µs/channel.

Note: Inputs to these devices can be measured at the 5 µs/channel rate. However, some

DBK expansion options require more settling time and can be measured at the

slower rate of 10 µs/channel or 1000 µs/channel. All types of measurements can

be mixed within the scan group.

Synchronization. Each device includes two SYNC ports. Up to four units can be

synchronized, with respect to each other. Synchronization is accomplished by connecting one

DaqBook device SYNC port to the SYNC port of a second DaqBook device. A unit can be
scan-synchronized

and triggered from any other DaqBook that is connected to it via a SYNC

cable. Either one of a unit’s two sync ports, or both, can be used. Note that the total

combined length of the Sync cables is not to exceed 15 feet (4.57 m). Chapter 8 includes a

block diagram regarding synchronizing multiple units.