beautypg.com

Measurement Computing DaqBook 2000 Series User Manual

Page 67

background image

SYNC Notes:

When a DaqBook is programmed as a synchronization slave, the TTL trigger source is

automatically derived from the SYNC port.

When a DaqBook is programmed as a synchronization master, it will output its trigger

condition on the SYNC port.

When synchronizing two or more DaqBooks, the slave units should have at least 0.1 µs of
dead time

in the scan period. By “dead time,” we mean a duration in which no channels

are sampled. This accommodates fundamental differences in the clocks [for each

DaqBook].

All DaqBooks that are connected via SYNC cables can be scan-synchronized to within

0.1 µs of each other.

A maximum of 4 units can be synchronized; scan-synchronous (post trigger).

SYNC cables are not to exceed a total combined length of 15 feet (4.57 m).

Synchronous Input Operations

DaqBooks allow synchronous scanning and acquisition of Analog Input, Digital Input and Counter Input

Data at up to 200 kHz aggregate scanning rates. The Analog Input data can be either main unit or

expansion modules from P1 compatible analog input modules. The Digital Input data can be main unit 8-

bit P2 (8255) digital input or P2 compatible DBK digital input expansion modules. For DaqBook/2001,

DaqBook/2005 and the DaqOEM equivalents, 16-bit P3 digital inputs are applicable. DaqBook/2020 has

no P3 connector, but uses of BNC connectors for the same functionality.

Analog Input Channels

P1 – Channels 0 through 15

DaqBooks allow analog input configuration of themselves, as well as for the P1 compatible DBK analog

input expansion modules.

Channel Selection and Mode Settings

The DaqBook/2001, /2005 and DaqOEM/2001 and /2005 can accept up to 16 single ended or up to 8

differential-ended inputs and can be programmed for single-ended or differential-ended on a per channel

basis. Just one analog channel is sacrificed when a DBK expansion module is enabled. See DBK

documentation in the DBK Option Cards & Modules User’s Manual (p/n 457-0905) for additional

information.
DaqBook/2020 can accept up to 14 single-ended inputs. One analog channel is sacrificed when a DBK

expansion module is enabled. See DBK documentation in the DBK Option Cards & Modules User’s
Manual

(p/n 457-0905) for additional information.

Channel Range and Polarity

DaqBook channels can be programmed for either unipolar or bipolar mode with gain settings

of 1, 2, 4, 8, 16, 32 and 64.

Channel Sampling Interval

DaqBooks allow for programmable sampling intervals of 5 us, 10 us, or 1000 us on a per channel basis.

This mode allows some DBK expansion channels [which change slowly but offer a higher degree of

accuracy] to be sampled at a longer interval, while channels that change more rapidly to be sampled using a

shorter interval. Each 5 µs, 10 µs, or 1000 µs interval reduces the maximum aggregate acquisition rate for

the entire scan by that amount.