CB-7011/CB-7011D, CB-7018 User’s Manual

2.20 ~AAEV

Description: Enable/Disable Calibration

Syntax: ~AAEV[CHK](cr)

~

delimiter character

AA

address of setting module (00 to FF)

E

command for enable/disable calibration

V

1=enable calibration, 0=disable calibration

Response: Valid Command:

!AA[CHK](cr)

Invalid Command:

?AA[CHK](cr)

Syntax error or communication error may get no

response.

!

delimiter for valid command

?

delimiter for invalid command

AA

address of response module(00 to FF)

Example:

Command: $010

Receive: ?01

Perform address 01 span calibration, return it is not ready for

calibration.

Command: ~01E1

Receive: !01

Set address 01 to enable calibration, return success.

Command: $010

Receive: !01

Preform address 01 span calibration, return success.

Related Command:

$AA0 (Sec. 2.5), $AA1 (Sec. 2.6)

Related Topic:

Calibration (Sec. 1.9)