2.6 $AA1
Description: Perform Zero Calibration
Syntax: $AA1[CHK](cr)
$
delimiter character
AA
address of setting module (00 to FF)
1
command for performing zero calibration
Response:
Valid Command:
!AA[CHK](cr)
Invalid Command:
?AA[CHK](cr)
Syntax error or communication error may get no re-
sponse.
!
delimiter for valid command
?
delimiter for invalid command or the calibration is not
enabled
AA
address of response module(00 to FF)
Example:
Command: $011
Receive: !01
Perform address 01 zero calibration, return success.
Command: $021
Receive: ?02
Perform address 02 zero calibration, return the calibration is
not enabled before perform calibration command.
Related Command:
$AA0 (Sec. 2.5), ~AAEV (Sec. 2.20)
Related Topics:
Calibration (Sec. 1.9)