10 setting up the ethernet interface – Guralp Systems CMG-6TD User Manual
Page 32

Installing the 6TD
The illustration above shows serial data in blue, network data in yellow and
power connections in red.
The Lantronix module can be powered down with the CD24 command
ETHER DISABLE
and powered up with the command
ETHER ENABLE
Certain operations, such as a firmware upgrade, can result in power to the
Lantronix module being turned off. In these cases, the ETHER ENABLE
command should be used to restore power.
4.10 Setting up the Ethernet interface
CMG-6TD instruments with Ethernet features installed use an embedded
Lantronix Wi-Port-NR module to provide the network interface. This module
can be configured using a built-in Web server.
Before you can access the Web server, however, you will need to assign the
device an IP address. This can be done using Lantronix' DeviceInstaller
utility for Microsoft Windows, or using a DHCP server. You will need a PC
with a network interface installed.
32
Issue F - February 2014