17 Insertion Gain

Insertion gain is the difference between aided and unaided ear canal SPL (REIG = REAG - REUG). It inherently
assumes a sound-field audiogram. For non-linear hearing aids, REIG is stimulus-dependent and the dependency
is specific to each hearing aid. Insertion gain fitting methods ignore these facts. Viewport or Speechmap is
recommended for fitting all linear and non-linear hearing instruments.

Refer to On-Ear Measures - Setup and On-Ear Measures Screen Setup for terminology and information on
preparing the system for use and for proper positioning of the client and the probe tube.

Insertion gain in SPL

1. Press , then highlight and [On-ear measures], then highlight and [Insertion gain].

2. Highlight and [Instrument] to chose from BTE, ITE, ITC, CIC, CROS, Body. This changes the

CORFIG used for 2cc targets.

3. Highlight and fitting method to chose from FIG6, NAL-NL1, NAL-R, POGO II, Berger, Libby,

User, None