Test box directional procedure – Audioscan Verifit User Manual
Page 56

Verifit
®
User's Guide Version 3.12
May 2015
Test box directional procedure
To test directional function in the test chamber:
1. Mount the hearing aid to the appropriate coupler (see Coupling the hearing instrument).
2. Position the hearing instrument in the test chamber as shown in 9.6: Directional ITE positioning or 9.7:
Directional BTE positioning.
3. Press 
4. Highlight and 
appear on the screen - the heavier one is the response to the left speaker signal and the lighter one the 
response to the right speaker signal. For a non-directional instrument, the two curves will be nearly identical. 
Setting the instrument for directional operation should produce a separation of the curves. 
5. On the Test control poster, highlight and 
from the drop-down list. For an adaptive directional instrument, a stimulus level between 60 and 70 dB 
should cause the separation to change over a 15 – 45 second period. For a noise-reducing instrument, both 
curves may change with time and stimulus SPL. 
6. Highlight and 
list box (defaults to 0 dB). A S/N ratio between 0 and 6 dB will usually induce directional performance after 
15 – 45 seconds. Optionally, highlight and 
alternating between the left and right test speakers. Speech-adaptive directional instruments may appear to be 
non-directional if [Speech S/N] is set to Off. 
56
