Calibration check for probe module – Audioscan Verifit User Manual

Page 65

background image

Verifit

®

User's Guide Version 3.12

May 2015

+/- 5 dB.

6. To change the calibration interval, highlight & [Daily] or [Weekly].

Press and repeat the process for the other probe module.

Calibration check for probe module

The calibration curve for the probe module is a comparison of the response of the probe microphone (with tube)
to the response of the on-ear reference microphone. Any change in the curve indicates a change in one of these
microphones (or the probe tube). See Calibration of on-ear probe microphone. As part of a regular calibration
schedule, or to investigate changing calibration curves, the on-ear reference microphone may be checked against
a high quality sound level meter.

1. Press , then highlight and [On-ear measures], then highlight and [Manual control].

2. Highlight & [Level], then highlight & 70 from the drop-down list.

3. Highlight and [Frequency], then highlight and 2000 Hz from the drop-down list.

4. Highlight, then [Start test] (which changes to [Stop test]).

5. Hold the probe module about 18 in. (0.5 m) in front of the sound-field speaker.

6. Hold the microphone of a calibrated sound level meter next to the on-ear reference microphone as shown.

Set the sound level meter to C weighting and slow.

7. The sound level meter should read 70 ±2 dB.

8. Repeat with the frequency at 250 Hz.

9. To stop the measurement, highlight and [Stop test].