Sound level meter using manual control – Audioscan Verifit User Manual
Page 60

Verifit
®
User's Guide Version 3.12
May 2015
Sound level meter using manual control
When set for a stimulus level of 0 dB (off), the Manual control mode allows the Verifit to be used as a sound
level meter. Sound at the test box reference microphone and the coupler microphone may be analyzed by 1/12th
octave, A-weighted or C-weighted filters. A and C-weighted measurements are limited to frequencies above 100
Hz and the averaging time and update rate are 384 ms.
1. Press
2. Highlight and
down list.
3. Highlight and
weighted from the drop-down list.
4. For 1/12th octave filter, highlight and
desired frequency from the drop-down list.
5. Highlight and
measurement, highlight and
test.
NOTE: Microphone extension cables (VA-131 & VA-133) are available from Audioscan. Standard audio
extension cables should not be used. They will substantially increase noise levels.
60