Page 109

background image

Verifit

®

User's Guide Version 3.12

May 2015

3. Select Test 1 and record a response using [Pink noise] at Soft (55). It should be like a normal REUR. If not,

the ear may not be fully open.

4. Position the speaker at 45 degress to the poor ear.

5. Set Instrument to [CROS].

6. Select Test 2 and obtain a response using [Pink noise] at Soft (55). Adjust the response to match that

obtained in step 3.Position the speaker directly in front of the client. Select Test 3 and use [Pink noise] at
Soft (55) to check for a smooth response. Repeat with Instrument set to [BTE]. An irregular response may be
the result of reflections from nearby objects (including the operator) or it may indicate phasing problems in
the aids.

T-Loop fitting and verification in Speechmap

This test allows the T-Loop response of the hearing instrument to be measured so that it can be compared
against a 65 dB SPL acoustic response.

1. Set the instrument to 'T' (Telecoil input).

2. Position the hearing instrument on the 'T' inscription on the test chamber floor vertically oriented as worn by

a seated individual, as shown below.

3. Press 'Test' and select any stimulus type except MPO, Speech-live, or Pink Noise.