Ansi test loop test, Telecoil test results – Audioscan Verifit User Manual
Page 50
Verifit
®
User's Guide Version 3.12
May 2015
ANSI test loop test
ANSI S3.22 requires that the gain control be set to RTP (1996) or RTS (2003) before running this test. The
reference test gain (RTG) is required in order to calculate the STS (1996) or RTLS (2003). The easiest way to
meet both requirements is to run a [Linear] or [AGC] test before running this test. See ANSI 1996 Linear and
AGC tests; ANSI 2003 Linear and AGC tests
1. First, complete an ANSI [Linear] or [AGC] test.
2. Press
3. Highlight and
4. Set the aid to ‘T’ (telecoil input).
5. Highlight and
6. Position the hearing instrument on the ‘T’ inscription on the test chamber floor vertically oriented as worn by
a seated individual as shown.
7. Without moving the hearing instrument, press
(SPLIV) curve or press
If an RTG has been previously obtained, the TLS or RTLS will be shown.
Telecoil test results
50