4. Press
5. Highlight and
instrument is noise-adaptive, the curves will separate after 15 – 45 seconds. A speech-adaptive directional
instrument will remain omni-directional and no separation will be observed. Press
directional test result and end the test.
6. Press
Viewport Noise reduction test
1. On either the Viewport Test box or On-ear screen, highlight and
2. Leave the setup in the test box or on-ear as used for previous Viewport tests. (NOTE: The rear speaker is not
used for On-Ear noise reduction tests.)
3. Highlight and
output within the first second as a heavy curve. A real-time output curve (light) will continue to be displayed
and will drop as the noise reduction feature of the instrument takes effect. A measure of the overall reduction
in wideband RMS between the two curves is displayed as a noise reduction (NR) value to the right of the
curves
4. Press
5. Highlight and
NOTE: The order of noise stimulus presentations, the type of noise used and the level of the noise stimuli
can be configured through Viewport Setup in the Setup Menu. Hearing instruments with noise reduction can
vary in terms of their activation level. If an expected noise reduction result is not obtained at a given
presentation level, try increasing the level and repeating the test.