IEEE 488.2 and SCPI Command Operation
3.11.2 CALIBRATION SCPI Command Reference
In the following table a "C" in the SCPI column indicates that the command is SCPI-compliant.
The Q? column indicates whether the command can also be used for querying.
Command Description
SCPI
Q?
CALibrate
Calibration subsystem.
C
:DATA
Sets the values of the ten floating point
calibration constants:
1) not used
2) not used
3) output current DAC gain
4) output current DAC offset
5) output current protection DAC gain
6) output current protection DAC offset
7) voltage measurement ADC gain
8) voltage measurement ADC offset
9) current measurement ADC gain
10) current measurement ADC offset
Values are separated by space or comma.
N
:INITial
N
:CURRent
Sets the power-on default value of current.
N
Q
:PROTection
overcurrent protection.
N Q
:LOCK
Disables access to the non-volatile memory.
Prevents attempts to store calibration
values.
N
:MEASure
N
:CURRent
:ADC?
Returns the integer value of the A/D for the
current measurement.
N
:CALCulate
Calculates the value of the gain and offset
for current measurements.
N
:GAIN
Sets the value of the gain for current
measurements.
N Q
:OFFSet
Sets the value of the offset for current
measurements.
N Q
:POINt <1|2>
<0+NRf>
Sets the current measurement calibration
point (1 or 2). The actual output current is
measured with an external meter.
N
:VOLTage
N
:ADC?
Returns the integer value of the A/D for the
voltage measurement.
N
:CALCulate
Calculates the value of the gain and offset
for voltage measurements.
N
:GAIN
Sets the value of the gain for voltage
measurements.
N Q
:OFFSet
Sets the value of the offset for voltage
measurements.
N Q