beautypg.com

System scpi command subsystem, System scpi command subsystem -17, System scpi command summary -17 – AMETEK SFA Series Programming Manual User Manual

Page 29: System scpi command reference -17

background image

Sorensen SF Series

IEEE 488.2 and SCPI Command Operation

3.9

3.9.1

3.9.2

SYSTEM SCPI Command Subsystem

SYSTEM SCPI Command Summary

SYSTem
:ERRor?
:FAULt?
:VERsion?
:LOCAL

:LOCAL?
:LOCLOUT

SYSTEM SCPI Command Reference

In the following table a "C" in the SCPI column indicates that the command is SCPI-compliant.
The Q? column indicates whether the command can also be used for querying.

Command Description

SCPI

Q?

SYSTem

System subsystem.

C

:ERRor?

Queries Error Queue for next error/event entry
(first in, first out). Entries contain an error
number and descriptive text. A 0 return value
indicates no error occurred; negative numbers
are reserved by SCPI. The maximum return
string length is 255 characters. The queue
holds up to 10 error/entries. All entries are
cleared by the *CLS command.

C

:FAULt?

Returns four numeric values separated by
commas for the four system fault registers. See
System Fault Registers below. Valid response
is 128, 0, 0, 0 or 0, 0, 0, 0.

Response: , ,
,

N

:VERsion?

Returns a numeric value corresponding to the
SCPI version number for which the instrument
complies. The response is in the format
YYYY.V where the Y’s represent the year and
V represents the approved version number for
that year (e.g., 1995.0)

C

M550292-03 (Programming)

3-17