Page 10: Figure 2–1. sfa configuration switch -2, 2 rear panel configuration switch

Configuration
Sorensen SF Series
2.2 Rear Panel Configuration Switch
The DIP switch is accessible from the rear panel to allow configuration of the supply with the
installed DIA for the user’s particular system and application. The following figures show the
configuration, as set up in Section 2.1, and with GPIB address set to five (5) for the SFA.
Note: Two types of DIP switches are utilized; toggle and rocker. For toggle switches, the
shading indicates the position of the toggle switch. For rocker switches, the shading
indicates the depressed side.
Figure 2–1. SFA Configuration Switch
Figure 2–2. SFI Configuration Switch
2.2.1 Remote/Local
Selection
Set the rear panel Remote/Local switch to select remote or local operation.
REMOTE/LOCAL SWITCH
Switch Position
Description
ON
Remote operation selected. *
OFF
Local operation selected.
Front panel control is enabled. Unit will switch to remote
operation upon the first GPIB or RS-232 command.
*
In the ON position, the power hardware and GPIB card initialize to the remote state on
power turn on. In addition, front panel control remains disabled regardless of the state of
the GPIB interface REN line or the GTL command. The special command SYST:LOCAL
2-2 M550292-03
(Programming)