IEEE 488.2 and SCPI Command Operation
Sorensen SF Series
THE RAMP FUNCTION
The ramp function allows the user to transition from one current to another linearly in a specified
time period (100 ms - 99 sec with 100 ms programming resolution). A unit may ramp only
current.
Current ramping requires an appropriate resistive load.
SOURce[n]:STATus:BLOCK? "Status Flags" REGISTER
Bit Hex
Value Name
Description
0
0x001
remote mode
Remote mode was selected.
1
0x002
Not used.
2
0x004
PON SRQ
PON Service Request selected by rear panel switch.
3 0x008
SRQ
sent
GPIB Service Request issued and the serial poll has
not been received.
4
0x010
EEPROM
The EEPROM is locked.
5
0x020
trip occurred
Either an OV or OT trip occurred.
6
0x040
sense signal
Not used.
7
0x080
isolation signal Not used.
8
0x100
hold
Current output waiting for trigger.
9
0x200
fold
Foldback protection is enabled.
10
0x400
SRQ enabled
Service Request generation is enabled.
11
0x800
output
Output is on.
3.6
3.6.1
MEASURE SCPI Command Subsystem
MEASURE SCPI Command Summary
MEASure
:CURRent?
:CURRent
:AVErage
:AVErage?
:VOLTage?
:VOLTage
:AVErage
:AVErage?
3-12 M550292-03
(Programming)