Toggle mode – AMETEK PLA-PLW User Manual
Page 81

Transient
Level
Toggle
t
TOGGLE MODE:
In toggle mode, the eLOAD input alternates between two user-defined input levels. Unlike transient-continuous mode,
the input only changes from one level to another, one transition at a time.
Typical Application: Testing the transient recovery time for a CV power supply. The recovery time for a power supply is
defined as the time required for the supply to settle to a predefined settling band following a change in load current caused
by an induced transient (pulse).
LCL Operation:
1.
Using the mode hot keys, select operating mode (CC, CV, CR, CP)
2.
Set the MAIN level by holding down the encoder and rotate to change the value or enter numerically.
3.
Set the Transient level by pressing the “PEAK” hot key and edit the value using the encoder or enter manually.
4.
Select TOGGLE transient operation by pressing the “TOGGLE” hot key.
5.
Set the Rising and Falling Slew Time by pressing “SLEW,” rotate the encoder to select option, tap the encoder to
edit slew time value and press “ENTER” to confirm value entry.
6.
To begin the TOGGLE transient operation, press the “PULSE” hot key (pulse hot key will blink) and then press
the “INPUT” hot key. To activate the TOGGLE, press the “TRIGGER” hot key.
Main
Level
3-27