AMETEK PLA-PLW User Manual
Page 104

CC0
LO= CNT: Input
measured
voltage from Curr.DMM, then press ENTER.
CC0
HI = CNT: Input
measured
voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
Press “
3” to select MAIN CC1. Then press enter.
CC1 LO= CNT: Input
measured
voltage from Curr.DMM, then press ENTER
CC1
HI = CNT: Input
measured
voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
Press “
3” to select MAIN CC2. Then press enter.
CC2 LO= CNT: Input
measured
voltage from Curr.DMM, then press ENTER
CC2
HI = CNT: Input
measured
voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR0) LOW RANGE MAIN LEVEL CALIBRATION
CR low range main level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of load.
Use “
4” and “3” until you see MAIN CR0 parameter. Press enter:
CR0
LO= CNT:
Input measured voltage from Volt.DMM, then press ENTER.
CR0
LO= CNT:
Input measured voltage from Curr.DMM, then press ENTER.
CR0
HI= CNT:
Input measured voltage from Volt.DMM, then press ENTER.
CR0
HI= CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR1) MIDDLE MAIN LEVEL CALIBRATION
CR middle range main level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.
Use “
4” and “3” until you see MAIN CR1 parameter. Press enter:
CR1
LO= CNT:
Input measured voltage from Volt.DMM, then press ENTER.
CR1
LO= CNT:
Input measured voltage from Curr.DMM, then press ENTER.
CR1
HI= CNT:
Input measured voltage from Volt.DMM, then press ENTER.
CR1
HI= CNT:
Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message
should be display on the front panel LCD.
CONSTANT RESISTANCE (CR2) HIGH RANGE MAIN LEVEL CALIBRATION
CR high range main level calibration requires a power supply that has high voltage and medium current, but operator has
to estimate the power consumption of load.
Use “
4” and “3” until you see MAIN CR2 parameter. Press enter:
CR1
LO= CNT:
Input measured voltage from Volt.DMM, then press ENTER.
5-4