AMETEK PLA-PLW User Manual

Page 104

background image

CC0

LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER.

CC0

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

Press “

3” to select MAIN CC1. Then press enter.

CC1 LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER

CC1

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

Press “

3” to select MAIN CC2. Then press enter.

CC2 LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER

CC2

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CONSTANT RESISTANCE (CR0) LOW RANGE MAIN LEVEL CALIBRATION

CR low range main level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of load.

Use “

4” and “3” until you see MAIN CR0 parameter. Press enter:

CR0

LO= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR0

LO= CNT:

Input measured voltage from Curr.DMM, then press ENTER.

CR0

HI= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR0

HI= CNT:

Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CONSTANT RESISTANCE (CR1) MIDDLE MAIN LEVEL CALIBRATION

CR middle range main level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.

Use “

4” and “3” until you see MAIN CR1 parameter. Press enter:

CR1

LO= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR1

LO= CNT:

Input measured voltage from Curr.DMM, then press ENTER.

CR1

HI= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR1

HI= CNT:

Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CONSTANT RESISTANCE (CR2) HIGH RANGE MAIN LEVEL CALIBRATION

CR high range main level calibration requires a power supply that has high voltage and medium current, but operator has
to estimate the power consumption of load.

Use “

4” and “3” until you see MAIN CR2 parameter. Press enter:

CR1

LO= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

5-4