6. PANEL LOCK
7. ETHERNET-------press ENTER key, use

43 to select ADRESS/NETMASK/GATEWAY; press ENTER.

E: Authenticate Menu

- press ENTER key:

*Use

43 to select the next menu setting then press ENTER key to enter the selected menu.

* (Use CLEAR key to back space or to correct the wrong value).

1. SERIAL NO.

----------press ENTER key. Serial NO. should start with 000B87xxxxxx.

2. INTERFACE------

press ENTER key. GPIB/ETHER/USB should be display according to customer’s order.

3. CODE--------------

press ENTER key, Input the code from AMREL, then press ENTER.

LOAD CONTROL MENU

STEP PARAMETERS

Selects the number of Step Count (0 ~ 65535 & infinity) and different Step Modes (STEP, STEP AUTO, and STEP ONCE)

MEASURE SENSE

This feature allows selection of local sense or remote sense of voltage. Remote sense measures voltage at a different
sense point than the bus bars to prevent large voltage drops over lengthy wires. If the voltage reading between the remote
sense and bus bar differ by greater than a set threshold, Remote Sense Fault (RSF) will trigger and shutdown the input.

Note: The eLOAD’s voltage measuring points (positive and negative) can be configured to two locations. One is from the
power input terminals called Local; the other is from the remote sense terminal block in the rear panel called Remote
(Remote option is only available in certain models). The purpose of the Remote sense is for more accurate voltage
measurement when longer cables are used (large voltage drop caused by long cables).

PORT 0

A general purpose output port that generates a TTL compatible signal; active high (5V) when port is on and inactive low
(0V) when port is turned off. The PORT 0 feature can be used to control external devices such as a DC Contactor or a
relay. To enable PORT 0, simply follow LOAD CONTROL

Æ PORT 0 Æ ON.

EXTERNAL CONTROL

The external control function allows the eLOAD’s maximum range (L/M/H) voltage, current, power and resistance to be
scaled from 0 to 10 Volts. The external voltage signal will control the eLOAD to operate at the corresponding
current/voltage/resistance/power. Please see the table below for further details.

Note: AMREL eLOAD’s external analog programming accepts bipolar Vac < 10Vpp.

EXT_CTRL Voltage

Current Range

Range Current Rating

Main Curent Value

0 V

L

6 A

0 A

5 V

L

6 A

3 A

10 V

L

6 A

6 A

0 V

M

30 A

0 A

5 V

M

30 A

15 A

10 V

M

30 A

30 A

0 V

H

60 A

0 A

5 V

H

60 A

30 A

10 V

H

60 A

60 A

1: SERIAL NO.

432: INTERFACE 43 3: CODE

3-7