beautypg.com

Osc range selection – AMETEK PLA-PLW User Manual

Page 127

background image

Command Syntax

STEP:CURRent:STATe

Parameters

0 | OFF(0) | 1 | ON(1) | 2 | AUTO(2) | 3 | ONCE(3)

Query Syntax

STEP:CURRent:STATe?

Returned Parameters

0=OFF

1=ON

2=AUTO

3=ONCE

NOTE: Please refer to the eLOAD Programming manual and Operation Manual for further Step
Operation and comprehensive Step SCPI Commands.

5. OSC Range Selection

The OSC Range selection now includes DISABLE + DEFAULT, DISABLE + OSC1,
DISABLE + OSC2 & DISABLE + OSC3. With the new selections, the eLOAD provides
more bandwidth selections for applications with dc load + ac perturbation. The
DISABLE OSC range will stop detection of oscillating conditions; in the case ac + dc
inputs are required for testing applications.

Local Control

MENU

LOAD CONTROL

Î OSCILLATE Î OSC CONTROL

Î DEFAULT

Î OSC1

Î OSC2

Î OSC3

Î DEFAULT + DISABLE
Î OSC1 + DISABLE
Î OSC2 + DISABLE
Î OSC3 + DISABLE

Remote Programming

SYSTem:OSCillate:PROTection

This command selects the Bandwidth (selects the programmable slew rate range) of the eLOAD
and enables/disables Oscillation Protection.

Command Syntax

SYST:OSC:PROT

Parameters

Mode>

0: DEFAULT
1: OSC1

2: OSC2

3: OSC3

4: DEFAULT + disabled

5: OSC1 + disabled

6: OSC2 + disabled

7: OSC3 + disabled

Query Syntax

SYST:OSC:PROT?

Returned Parameters

0 ~ 7

C-7