Run command: “SYST:OFFS? ” to get calibration offset values where “mode” is from 0 to 39. So
there should be 40 values for offset data.

2.

Once you obtain all the calibration information, you can resave this calibration information to the electronic load.

Resave calibration slope date to electronic load.

Run command: “SYST: SLOPE where “mode” is from 0 to 39, and the value is the
corresponding slope value obtained from 1.

Resave calibration offset date to electronic load.

Run command: “SYST: OFFS where “mode” is from 0 to 39, and the value is the
corresponding offset value obtained from 1.

After re-saving the calibration information for electronic load, you need test the CV and CC mode,

and check if the reading back values are within spec.

5-9