Page 106

background image

CV2 LO= CNT: Input

measured

voltage from Volt.DMM, then press ENTER

CV2

HI = CNT: Input

measured

voltage from Volt.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CC LOW/MIDDLE/HIGH RANGE, TRANSIENT LEVEL CALIBRATION

CC transient level calibration requires a power supply that covers full range of high range input current of load. Estimate
the power consumption of the electronic load, and adjust the voltage input to a safe limit for the load.

Use “

4” and “3” until you see TRAN CC0 parameter. Press enter:

CC0

LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER.

CC0

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

Press “

3” to select MAIN CC1. Then press enter.

CC1 LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER

CC1

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

Press “

3” to select MAIN CC2. Then press enter.

CC2 LO= CNT: Input

measured

voltage from Curr.DMM, then press ENTER

CC2

HI = CNT: Input

measured

voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CR LOW RANGE TRANSIENT LEVEL CALIBRATION

CR low range transient level calibration requires a power supply that has low voltage and high current, but operator has to
estimate the power consumption of electronic load.

Use “

4” and “3” until you see TRAN CR0 parameter. Press enter:

CR0

LO= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR0

LO= CNT:

Input measured voltage from Curr.DMM, then press ENTER.

CR0

HI= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR0

HI= CNT:

Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD.

CR MIDDLE RANGE TRANSIENT LEVEL CALIBRATION

CR middle range transient level calibration requires a power supply that has medium voltage and medium current, but
operator has to estimate the power consumption of electronic load.

Use “

4” and “3” until you see TRAN CR1 parameter. Press enter

CR1

LO= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR1

LO= CNT:

Input measured voltage from Curr.DMM, then press ENTER.

CR1

HI= CNT:

Input measured voltage from Volt.DMM, then press ENTER.

CR1

HI= CNT:

Input measured voltage from Curr.DMM then press ENTER key. A “SUCCESS” message

should be display on the front panel LCD..