Casella CEL CEL-368 User Manual
Page 86

Measurement ranges - CEL-368:
Range W'ghting Display
Primary
Primary
Pulse
Setting
Range
Linearity
Linearity
Range
Range*
Range
Pk values
IEC 651
IEC 804
IEC 804
10-80
A
10-80.1 dB
23-66 dB
23-80 dB
60 dB
C
10-80.1 dB
26-66 dB
26-80 dB
57 dB
Linear
10-80.1 dB
32-66 dB
32-80 dB
51 dB
40-110 A
40-110.1 dB 40-96 dB
40-110 dB 73 dB
C
40-110.1 dB 40-96 dB
40-110 dB 73 dB
Linear
40-110.1 dB 40-96 dB
40-110 dB 73 dB
70-140 A
70-140.1 dB 70-126 dB 70-140 dB 73 dB
C
70-140.1 dB 70-126 dB 70-140 dB 73 dB
Linear
70-140.1 dB 70-126 dB 70-140 dB 73 dB
Note*
Lower reading is quoted at +10 dB on noise floor.
Upper reading allows crest factor 5.
Dynamic range:
70 dB.
Accuracy: (under reference conditions)
CEL-268: 0.7 dB,
CEL-368: 1.0 dB.
Typical instrument frequency range (within +0.5 to -3 dB)
CEL-268 - Type 1:
70-140 range: 5 Hz to 63 kHz,
40-110 range: 5 Hz to 63 kHz,
10-80 range:
5 Hz to 28 kHz.
CEL-268/368 Handbook