Specification – Casella CEL CEL-368 User Manual
Page 85

9.
SPECIFICATION
CEL-268 - Type 1:
Precision Sound Level Meter according to IEC 651: 1979, Type 1,
Precision Integrating Sound Level Meter according to IEC 804:
1985, Type 1,
Precision Sound Level Meter (with "R" type random incidence
microphone) according to ANSI: S1.4-1983, Type S1.
CEL-368 - Type 2:
Sound Level Meter according to IEC 651: 1979, Type 2,
Integrating Sound Level Meter according to IEC 804: 1985,
Type 2,
Sound Level Meter according to ANSI: S1.4-1983, Type S(1).
Measurement ranges - CEL-268:
Range W'ghting Display
Primary
Primary
Pulse
Setting
Range
Linearity
Linearity
Range
Range*
Range
Pk values
IEC 651
IEC 804
IEC 804
10-80
A
10-80.1 dB
26-60 dB
26-80 dB
57 dB
C
10-80.1 dB
33-60 dB
33-80 dB
50 dB
Linear
10-80.1 dB
45-60 dB
45-80 dB
38 dB
40-110 A
40-110.1 dB 40-90 dB
40-110 dB 73 dB
C
40-110.1 dB 40-90 dB
40-110 dB 73 dB
Linear
40-110.1 dB 40-90 dB
40-110 dB 73 dB
70-140 A
70-140.1 dB 70-120 dB 70-140 dB 73 dB
C
70-140.1 dB 70-120 dB 70-140 dB 73 dB
Linear
70-140.1 dB 70-120 dB 70-140 dB 73 dB
Note*
Lower reading is quoted at +10 dB on noise floor.
Upper reading allows crest factor 10.
CEL-268/368 Handbook
Page - 84