BINDER KBF LQC 240 User Manual
Page 68
KBF LQC (E5.3) 07/2014
page 68/114
(continuation)
Notifying sequence
1
2
3
Group Event
Note
(blue field)
Alarm
(red field)
Zero-voltage
relay alarm
outputs (option)
3
X = representing the actual val-
ues of VIS and UVA
LIGHT kLUX W/m
2
X = representing the integrated
values of VIS and UVA
DOSES MLUXh
Wh/m
2
Dose target value LUX reached
LUX DOSE
REACHED
immediately
UVA AND LUX
DOSE REACHED
when also dose target
value UVA is reached
Dose target value UVA reached
UVA DOSE
REACHED
immediately
UVA AND LUX
DOSE REACHED
when also dose target
value LUX is reached
Switch “LIGHT INTEGRATOR” is
in position “OFF”
LIGHT INTEGR. OFF
4
Maximum temperature of 60°C
with light sensor is reached (only
if at least 1 light sensor is
plugged-in)
LIGHT SENSOR 60
°C!
immediately
5
Temperature deviation of more
than +/- 2 °C of the entered set-
point
TEMP RANGE
immediately
TEMP RANGE
after 16 Min.
after 16 min.
Humidity deviation of more than
+/- 5 % r.H. of the entered set-
point
HUMID RANGE
immediately
HUMID RANGE
after 16 min.
after 16 min.
6
Humidity system turned off with
switch (4)
HUMID OFF
immediately
Door open
DOOR OPEN
immediately
DOOR OPEN
after 5 min.
7
With option locking of the keyboard (BINDER Individual, chap. 17.7):
Locked keyboard
KEY LOCK
immediately
Power failure
immediately
The indicated intervals refer to the time after occurrence of the error or notified condition.
Notifying messages from one group (e.g. Group 2) mask the notifying messages of the subsequent
Group (e.g. Group 3). The same is true for the alarm messages of different groups.
Notifying, and alarm messages of Group 3 are displayed in intervals, so messages of subsequent groups
can appear for a moment alternately.