6.
Settings of the MB1 controller
6.1 Selection of the MB1 controller’s menu language
The MB1 display program controller controls the temperature and humidity inside the constant climate
chamber. In addition to displaying the actual values, it permits cumulative measurement of the light doses
of UVA and the visible spectral range.
The controller communicates by a menu guide using real words in German, English and French. The
selection of the desired menu language is located in the sub-menu “User-Level” of the “User-Settings”
menu. Select menu point “Language“.
User level code no.
1 (factory setting)
The row of buttons below the display is context-sensitive. The inscription above the buttons on the display
defines the button’s function.
Do NOT change the temperature unit from °C to °F.
User Level
Date and time
Summer time
Language
Temperature unit
Buzzer
Safety controller
User-code No.
English
°C
1
Active
Sollwert Art
Offset
Language
English
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 36.8 °C
Safety control. Set. 38.5
°C
29
User-Settings
08:43:55 15.12.13
Configuration 2
Configuration 1
Parameters
Choose variation
08:43:55 15.12.13
Configuration 2
Parameters
Choose variation
Configuration 1
User-Settings
CONFIG
4 X