4.2 Wastewater connection
Fasten the wastewater hose to the waste-water connection “OUT” (17) on the rear of the unit (olive ∅ 14
mm). Observe the following points:
•
You can use a part of the delivered water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against at least 95 °C / 203°F.
•
Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of
3 m.
•
Protect both ends of the drainage hose with two of the four delivered hose clamps.
Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off
only when required, thus there is no continuous waste water flow.
(22)
(16)
(18) (17)
(16) Power cable
(17) Waste-water connection “OUT” with hose
olive for hose ½“
(18) Freshwater connection “IN” with screw
thread ¾’’ for hose ½“, with union nut
(22) Socket for optional freshwater can (chap.
Figure 10: Rear view KBF LQC with water connections
Protect the waste water supply at both sides with the delivered hose clamps.