3 storage, Caution – BINDER KBF LQC 240 User Manual
Page 23
![background image](https://www.manualsdir.com/files/767129/content/doc023.png)
KBF LQC (E5.3) 07/2014
page 23/114
You can order transport packing for transportation purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the steam humidifying system has NOT been emptied: +3 °C / 37.4°F to +60 °C / 140°F.
•
After BINDER Service has emptied the steam humidifying system: -10 °C / 14°F to +60 °C / 140°F.
With temperatures below +3 °C / 37.4°F, water must be completely removed from the humidifying sys-
tem.
CAUTION
Transport below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.
Contact BINDER Service before any transportation below +3 °C / 37.4°F.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-
rary decommissioning (chap. 19.2).
Permissible ambient temperature range during storage:
•
If the steam humidifying system has NOT been emptied: +3 °C / 37.4°F to +60 °C / 140°F.
•
After BINDER Service has emptied the steam humidifying system: -10 °C / 14°F to +60 °C / 140°F
With temperatures below +3 °C / 37.4°F, water must be completely removed from the humidifying sys-
tem.
CAUTION
Storage below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.
Contact BINDER Service before any transportation below +3 °C / 37.4°F.
Permissible ambient humidity: max. 70 % r.H., non-condensing
CAUTION
Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H.
for a long period.
Dry the appliance completely before shut-down:
•
Set the humidity to 0 % r.H. and turn on the humidity switch (4).
•
Set the temperature set point to 60 °C / 140°F for approx. 2 hours (Manual mode).
•
Only then, shut down the unit at the main power switch (3) and close the tap of the
water supply.
When after storage in a cold location you transfer the unit to its warmer installation site, condensation
may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and
is completely dry.
In case of a prolonged temporal decommissioning: Leave the unit door open or remove the access port
plugs.