beautypg.com

3 menu settings in the “user-settings” menu – BINDER KBF LQC 240 User Manual

Page 38

background image

KBF LQC (E5.3) 07/2014

page 38/114

EVENT LIST

Overview over the last 16 events or error occurrenc-

es of the unit.

CHART RECORDER FUNCTION

Graphical display of the current temperature and

humidity, and light data, and review of the previous

measurements on a historical display. A memory

interval of 5s corresponds to a supervision period

of 2.5 days

6.3 Menu settings in the “User-settings” menu

User-settings

Instrument data

Contrast

Displ. Power down

Contin. operation

User Level

Safety control. Act + 39.8 °C

Safety control. Set. + 42.0 °C

29

Instrument data

Instrument Name
Enter an individual name of the constant climate chamber.
Address
Enter a controller address (1 to 30) for operation with the communication

software APT-COM™.
All other entries are relevant only for service purposes.

Contrast

No function.

Displ. power down

Switch off event
Do not change the entry “Wait. Period”.
Waiting period
You can enter a delay time after which the display, following manual activa-

tion, will automatically be turned off again, on condition that the moment is

outside the operation time defined in menu ”Contin. operation”.

Contin. operation

Enter an operation time to determine the period of display activity. Outside

the defined period, the display is automatically turned off. Pressing down any

key will reactivate the display. After the time set in menu “Displ. power down“,

the display will turn off again as far as the actual time is not within the opera-

tion time fixed in menu “Cont. operation“.

User Level

Toggle here to the display menu “User Level” (chap. 6.4) by entering a pass-

word. Factory default setting for this password is +00001. You can change

the password (“user code”) in the menu “User Level”.

08:43:55 15.12.13

08:43:55 15.12.13

This manual is related to the following products: