BINDER KBF LQC 240 User Manual
Page 53

KBF LQC (E5.3) 07/2014
page 53/114
10.5 Program entry as set-point ramp or as set-point step
In order to avoid incorrect programming, we recommend plotting the temperature profile and the humidity
profile (chart templates in chap. 10.11 and 10.12) and entering the values into a table (templates in chap
10.13 and 10.14).
The controller provides 8 operation lines (non-functional with standard unit) that can be activated or de-
activated for each program section.
Program entry as set-point ramp (example of a temperature program)
W/°C
t/min.
01
02
03
04
0
20
40
60
80
100
30
120
180
380
05
Tolerance
maximum
Tolerance
maximum
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan Section time Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01
40
100 % 00:30:00
Off
1
0
-1999
+9999
02
60
100 % 01:30:00
Off
1
0
-5
+5
03
90
100 % 01:00:00
Off
1
0
-2
+2
04
90
100 % 03:20:00
Off
1
0
-1999
+9999
05
20
100 % 00:00:01
Off
1
0
-1999
+9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the MB1 program controller:
08:43:55 15.12.13