page 96/107
Unit size
23
53
115
Temperature data
Temperature range approx. 15 °C / 27 °F
above room temperature up to
°C / °F
200 / 392
200 / 392
200 / 392
Temperature fluctuation
± K
0.1
0.1
0.1
Temperature uniformity
(variation)
1)
at 100 °C / 212°F
± K
1.5
2
3,5
at 200 °C / 392°F
± K
3
4.5
9
Heating up time
1) 2)
to 100 °C / 212°F
min
65
80
95
to 200 °C / 392°F
min
100
115
150
Max. temperature of inner chamber surfac-
es
°C / °F
320 / 608
320 / 608
320 / 608
Vacuum data
Vacuum connection with small flange
DN mm / inch
16 / 0.63
16 / 0.63
16 / 0.63
Measuring access port with small flange
DN mm / inch
16 / 0.63
16 / 0.63
16 / 0.63
Inert gas connection with flow-limiter
Adapter with hose olive
∅
mm / inch
8 / 0.31
8 / 0.31
8 / 0.31
Permitted end vacuum
mbar / inHg
1x10
-2
/
0.0003
1x10
-2
/
0.0003
1x10
-2
/
0.0003
Leak rate
bar/h / inHg /h
1x10
-2
/
0.295
1x10
-2
/
0.295
1x10
-2
/
0.295
Classification of the explosion proof inner chamber
(according to ATEX Directive 94/9/EC)
II -3G IIB T3-T1 Gc X
Electrical data
IP protection type according to EN 60529
IP 54
IP 54
IP 54
Nominal voltage (±10%) 50/60 Hz
V
230
230
230
Nominal power
kW
0.80
1.20
1.90
Unit fuse 5 x 20 mm / 250V /
middle-time-lag (M)
Amp
10
10
10
Power plug
shock proof plug
Power plug vacuum module (option)
shock proof plug
Energy consumption
at 100 °C / 212°F
Wh/h
105
150
250
at 200 °C / 392°F
Wh/h
280
445
785
Over-voltage category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Legend:
1)
Values only in combination with aluminum racks
2)
Time to reach 98% of the set point temperature
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in ac-
cordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances
of 10% of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.