Figure 15: Possible temperature transitions
(with default setting “ramp” in the user level (chap. 10)
A heating up rate of 1.5 °C/min can be regarded as the realistic maximum.
If the oven is charged to full capacity, depending on the load, deviations from the specified
heating up times may occur.
Program entry as set-point ramp (example):
W/°C
t/min.
S01
S02
S03
S04
0
20
40
60
80
100
30
120
180
380
S05
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
60
01:30
S03
90
01:00
S04
90
03:20
S05
20
00:01
You can now enter the values of such a program table to the RD3 program controller (chap. 8.2).