Installation and connections, 1 vacuum expansion racks, Warning – BINDER VD 23 User Manual
Page 34: Caution
VDL (E2.1) 04/2014
page 34/107
WARNING
High concentration of inert gas.
Risk of death by suffocation.
∅
Do NOT set up units in non-ventilated recesses.
Ensure technical ventilation measures.
Respect the relevant regulations for handling these gases.
When decommissioning the vacuum drying oven, shut off the inert gas supply: Close
valve (6).
4.
Installation and connections
4.1 Vacuum expansion racks
The low-loss heat transfer to the material occurs via the pa-
tented aluminum vacuum expansion racks (also available in
stainless steel as an option). The strong tension causes the
racks to fit tightly against the interior wall and their large-
surface contact area ensures rapid and effective heat transfer.
The removable rack holders allow for easy positioning.
You can also remove the expansion racks for easy cleaning.
Do not remove them too often in order to prevent wear.
Figure 12: Using the expansion racks
•
Pushing the locking lever: The expansion rack is released and can be removed.
•
Pulling the locking lever: The expansion rack is pressed against the inner chamber walls.
Following each new tightening of an expansion rack, check that the lateral parts of the rack fit
closely over their whole surface to the inner chamber wall. This is necessary in order to ensure
the specified temperature exactitude.
CAUTION
Invalid calibration.
∅
Do NOT change between aluminum and stainless steel racks
Use the delivered expansion racks only