5 measuring access port vacuum 9 poles (option), Danger – BINDER VD 23 User Manual
Page 84

VDL (E2.1) 04/2014
page 84/107
15.5 Measuring access port vacuum 9 poles (option)
The measuring access port allows creating electrical connec-
tions for low voltages or sensors between the exterior and interi-
or of the VDL vacuum drying oven. A 9 poles plug for the outside
is included.
Figure 28: Measuring connection (15)
with measuring access port and supplied plug
Connections at the measuring access port
•
At the inner side of the measuring access port you can solder up to 9 cables. The inside connections
must be insulated against each other and against ground. Use 300 °C / 572°F solder.
•
The 9 contacts are conducted outside via the measuring access port. You can use them to connect a
device via the supplied plug.
Maximum load capacity of the switching contacts: 42 V AC/DC - 2A
DANGER
Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.
∅
Do NOT exceed the maximum switching load of 42 AC/DC – 2 Amp.
∅
Do NOT connect any devices with a higher load capacity.
Insulate the inside connections against each other and against ground. Use 300 °C /
572°F solder.
15.6 Communication software APT-COM™ 3 DataControlSystem (option)
The unit is regularly equipped with a serial interface RS 422 (9) that can connect the BINDER communica-
tion software APT-COM™ 3 DataControlSystem. The actual temperature value is given at adjustable in-
tervals. Programming can be performed graphically via PC. Up to 30 chambers with RS 422 interface can
be cross-linked. For further information, refer to the operating manual of the BINDER communication
software APT-COM™ 3.
Confirm that the interface mode is correctly set to “Modbus” in the user level (chap. 10).
Pin allocation of the RS 422 interface (9):
pin 2: RxD (+)
pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)
pin 7: GND