Danger – BINDER VD 23 User Manual
Page 44

VDL (E2.1) 04/2014
page 44/107
When changing the set-point, check the setting of the safety device (chap. 12).
The values entered in fixed-value entry mode remain valid after program run-off and are then
equilibrated.
If the week program timer is active, depending on the running week program another set-point (SP2) may
be targeted. Resulting high temperatures may exceed the maximum permissible drying temperature of a
solvent in the drying material. Therefore, deactivate the week program timer if you do not use it (default
setting, setting in the User level, chap. 10).
DANGER
Exceeding the ignition temperature of a solvent after the program ends.
Risk of fire and explosion.
Danger of death.
Deactivate the week program timer if you do not use it.