BINDER VD 23 User Manual
Page 64
VDL (E2.1) 04/2014
page 64/107
In the first step, select a program. This is on condition that a program has been entered previously (chap.
8.2) and that program type “2 programs with 10 sections each” has been selected in the user level (chap.
10).
Then define the settings for the program course. Two parameters can be set:
•
Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1
minute, and its maximum value is 99.59 (99 h 59 min). If the value is 00.00, the program will start im-
mediately.
•
Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be
entered. If the program is not going to be repeated, enter the value “0”. For indefinite repeats enter the
value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections.
In the last step start the selected program. These steps must be carried out subsequently.
Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before
starting a program.
Step 1 – Program selection (only with program type “2 programs” set)
Normal Display
Press program key
Display 1 shows
e.g. 1
(actual selection of the program)
Display 2 shows
SEL.PRG.
(select program 1 or 2)
Enter program number 1 or 2 with
arrow keys
Value is shown in display 1.
Next step – entry of program course settings
Press program key
Display 1 shows
e.g. 00.00
(entered delay time hh.mm)
Display 2 shows
RUN TIME
(enter delay time of program start)
Set delay time in hh.mm using arrow keys
Value is shown in display 1.
Press program key
Display 1 shows
e.g. -1
(actual selection of the number of program cycles)
Display 2 shows
REPEAT
(enter number of program cycles)
Select number of cycles -1, 0, 1 etc. using
arrow keys
Value is shown in display 1.