BINDER VD 23 User Manual
Page 56
VDL (E2.1) 04/2014
page 56/107
Program entry as set-point step (example):
W/°C
t/min.
S01 S02 S03 S04 S05 S06 S07 S08
0
20
40
60
80
100
30
120
180
380
Program table corresponding to the diagram (with default setting “Ramp”):
Section
SEC
Temperature
set-point
[ °C]
TEMP
Section length
[hh.mm]
TIME
S01
40
00:30
S02
40
00:01
S03
60
01:30
S04
60
00:01
S05
80
01:00
S06
80
00:01
S07
20
03:20
S08
20
00:01
You can now enter the values of such a program table to the RD3 program controller (chap. 8.2).
Program the end point of the desired cycle with an additional section (in our examples S05 for set-point
ramp and S08 for set-point step) with a section time of at least one minute. Otherwise, the program will
stop one section too early because the program line is incomplete.