Ntellisys, Daptive, Ressure – Nor-Cal Intellisys Adaptive Pressure Controllers User Manual

Page 8: Ontrollers, 2 - a/c powered buried box controllers interface

background image

Visit our Web Site

www.n-c.com

8

8

I

ntellIsys

A

dAptIve

p

ressure

C

ontrollers

APC-OP-LIT 1/12

Status LEDs

Valve

Open/Close

Switch

Momentary

Power

On/Off

Switch

3.2 - A/C Powered buried box Controllers Interface

Refer to the following diagrams and tables when connecting to an A/C powered APC Buried Box Controller including models APC-100-A, APC-200-A, APC-300-A, APC-

700-A and APC-800-A.

TAbLE 3.2.6 -

(J6) “APC-X00-A”ANALOG / TTL

CONNECTOR, Db-37 RECEPTACLE

PIn

FunctIon

1

N/C

2

N/C

3

N/C

4

Digital GND / RS232 Common

5

\CDG2 Select In\

6

\Position Control Select In\

7

N/C

8

\Close Valve In\

9

\Control Active Out\

10

\Analog Div 10 Select In\

11

\Analog SP Control Select In\

12

Digital GND / RS232 Common

13

TTL Out Common

14

N/C

15

N/C

16

N/C

17

N/C

18

Analog Valve Out (0-5 VDC)

19

\Valve Open Out\

20

N/C

21

N/C

22

N/C

23

\Valve Closed Out\

24

Fault Out

25

\Zero Gauge Select In\

26

See NOTE below

27

\Open Valve In\

28

PLO #2 Status

29

PLO #1 Status

30

N/C

31

N/C

32

N/C

33

Analog SP+ In

34

Analog SP- In

35

Analog GND

36

Analog CDG Out

37

Analog Valve Out

TAbLE 3.2.3 - (J5)

“APC-X00-A” VALVE

CONNECTOR,

Db-15 RECEPTACLE

PIn SIGnAl

ASSIGnment

1

Drive A+

2

Drive A-

3

Drive B-

4

Drive B+

5

Sense B-

6

Sense B+

7

Sense A-

8

Sense A+

9

Reserved

10

Reserved

11

Reserved

12

Reserved

13

Reserved

14

Reserved

15

Reserved

TAbLES 3.2.5 - (SW1) “APC-X00-A” DIP SWITCh

SETTINGS FOR SERIAL COmmUNICATIONS

BAuD rAte

SwItch 1

SwItch 2

19200

OFF (up)

OFF (up)

9600

ON (down)

OFF (up)

4800

OFF (up)

ON (down)

1200

ON (down)

ON (down)

PArIty

SwItch 3

SwItch 4

None

OFF (up)

Either ON or OFF

Odd

ON (down)

OFF (up)

Even

ON (down)

ON (down)

StoP BItS*

SwItch 5

1

OFF (up)

2

ON (down)

* If parity is enabled (either odd or even) then only 1 stop bit is

used and SW1 position 5 will be ignored

other

FunctIonS

SwItch

6

SwItch

7

SwItch

8

ON (down)

Enable

RS-485

Analog S.P.

0-5 VDC

Reserved

OFF (up)

RS-232

Analog S.P.

0-10 VDC

Reserved

TAbLE 3.2.1 - (J3 & J4)

“APC-X00-A” GAUGE

CONNECTOR Db-9 RECEPTACLE

(IDENTICAL PINOUTS)

PIn

SIGnAl

ASSIGnment

1

Signal input from gauge

2

N/C

3

N/C

4

+15VDC output to gauge

5

-15VDC output to gauge

6

N/C

7

N/C

8

Signal common from

gauge

9

Power common

output to gauge

TAbLE 3.2.4 - (J2) “APC-X00-A” RS-

232 SERIAL PORT, Db-9 RECEPTACLE

PIn

SIGnAl

ASSIGnment

1

RS485 (B)

2

RS-232 TX (data from

APC to host)

3

RS-232 RX (data from

host to APC)

4

N/C

5

Signal Common

6

N/C

7

N/C

8

N/C

9

RS485 (A)

TAbLE 3.2.2 - (J1) “APC-X00-A” A/C

POWER CONNECTOR

PIn

FunctIon

1

Neutral

2

Ground

3

Line in 100 to 240 VAC

1

3

2

VALVE

VALVE

POWER

TTL/ANALOG I/O

TTL/ANALOG I/O

GAUGE 1

GAUGE 1

SW1

SW1

!!

100-240V~
50-60 Hz

SERIAL

SERIAL

GAUGE 2

GAUGE 2

6.50

2.75

DEPTH: 10.60 ”

NOTE:

Function \names\ with a

backward slash indicate active low
input or output signals

NOTE:

For 100-, 200- &

300- series controllers,

pin 26 is defined as

\Hold Valve Select In\

NOTE:

For 700- &

800- series controllers,

pin 26 is defined as

\Valve Initialization

Enable In\