Ntellisys, Daptive, Ressure – Nor-Cal Intellisys Adaptive Pressure Controllers User Manual
Page 11: Ontrollers
11
Call toll free
800-824-4166
or 530-842-4457
•
FAX 530-842-9130
I
ntellIsys
A
dAptIve
p
ressure
C
ontrollers
APC-OP-LIT 1/12
2.51
5.74
5.69
4.78
Indicating LEDs
3.5 - DC Powered Controllers with Serial and DeviceNet I/O Interface
Refer to the following diagrams and tables when connecting to an D/C powered APC controller including models APC-100L-D, APC-200L-D, APC-300L-D, APC-700L-D and
APC-800L-D.
TAbLE 3.5.1 –
(J1) “APC-X00L-D”
DEVICENET NETWORk
CONNECTOR
PIn
FunctIon
1
Drain
2
Power In +
3
Power In -
4
CAN H
5
CAN L
1
4
3
2
5
TAbLE 3.5.2 – (J2) “APC-X00L-D”
AUXILIARy CONNECTOR,
Db-15 RECEPTACLE
PIn
FunctIon
1
Not used
2
RS-232 TX (data from IQ to
host)/485 (A)
3
RS-232 RX (data from host to
IQ)/485 (B)
4
Digital GND
5
Power RTN
6
Not used
7
Power +24V in
8
Not used
9
Analog CDG Out
10
Analog Valve Out
11
Analog GND
12
TTL Out Common
13
\Valve Open Out\
14
\Valve Closed Out\
15
Chassis GND
TAbLE 3.5.3 – (J3) “APC-X00L-D”
GAUGE CONNECTOR,
Db-9 RECEPTACLE
PIn #
FunctIon
1
Signal input from gauge 1
2
+15 V supply to gauge
3
-15 V supply to gauge
4
Not used
5
Signal input from gauge 2
6
Signal common from gauge 2
7
Not used
8
Signal common from gauge 1
9
Power supply common
TAbLE 3.5.4 - (J4) “APC-X00L-D”
VALVE CONNECTOR,
Db-15 RECEPTACLE
PIn
SIGnAl
ASSIGnment
1
Drive A+
2
Drive A-
3
Drive B-
4
Drive B+
5
Sense B-
6
Sense B+
7
Sense A-
8
Sense A+
9
Reserved
10
Reserved
11
Reserved
12
Reserved
13
Reserved
14
Reserved
15
Reserved