Nor-Cal Intellisys Isolation Pendulum Valve User Manual

80 psi cda or n, Input, Manual “open” button

background image

Intelli

sys

Isolation Pendulum Valve (IPV)

Quick-Start Guide

200mm, 250mm and 350mm Isolation Pendulum Gate Valves (IPV)

Nor-Cal Products, Inc. 1967 South Oregon St., Yreka, CA 96097 USA

LIT-IPV-QSG, 05/2006

Unpacking & Handling

Use care when unpacking an Intellisys Pendulum Valve.

Never lift by the Actuator or Motor, or set on edge.

Lift only on the body surfaces or lifting lugs.

Aluminum surfaces scratch easily.

A static-safe, particle-free

work surface is recommended. There is no need to remove parts during installation or inspection.
Removing or disassembling parts without authorization from a Nor-Cal rep will void the warranty.


Check The Package Contents:

Intellisys Isolation Pendulum Valve (IPV)

This Quick-Start Guide


Serial Label

A Serial Label on the bag shows the model number and serial number of your product. Additionally, Model,
Serial and Date Codes are engraved, and an inspection sticker is affixed to the side of the valve.





Warranty

Nor-Cal Intellisys Controllers and valves are warranted against defects and material workmanship for a period of
15 months after the date of shipment to the buyer. For complete warranty and Intellectual Property details,
please refer to the full manual.

Pre-Installation Functionality Check

Refer to the photograph below for locations of the items that follow.

WARNING: The valve plate movement presents a severe pinching/crushing
hazard. Please keep fingers, hands and other objects away from the valve
opening and associated moving parts.

Connect the Pneumatic Line
For pneumatically driven IPV valves, connect an unpressurized CDA or N

2

line to the input solenoid then slowly

bring up the pressure to 80PSI.

Open and Close Testing
To open the valve, manually depress and hold the small orange button on the solenoid near the CDA input.
Release the button to close the valve.

Connections & Interfaces


The IPV Pneumatic Air connector and
Manual Open/Close button are located
on the IPV Air Control Bracket, shown here:

Serial Label

Body Engraving

Inspection Sticker

¼” 80 PSI CDA

or N

2

input

(NOTE: 80-100 PSI is required

for proper valve sealing)

Manual

“Open” Button

Location of Air Control Bracket