6.6 ORP
• From GLP data main menu select the “ORP” option and press OK.
• All the information about the last ORP calibration appears: calibrated point,
time and date
• In case of restoring factory calibration data, the
display will show the warning message “Factory
calibr. restored”.
• Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NO
NO
NO
NO
NOTE
TE
TE
TE
TE
If no ORP calibration has been performed, a warning message advise
the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
6.7 TEMPERATURE
• From GLP data main menu select the “Temperature” option and press OK.
• All the information about the last temperature
calibration appears: calibrated point, time and
date.
• Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NO
NO
NO
NO
NOTE
TE
TE
TE
TE
If no temperature calibration has been performed, a warning message
advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.