Chapter 4 - CALIBRATION MODE
HI9828
HI9828
HI9828
HI9828
HI9828 allows to perform six different types of calibration, one for each param-
eter and also a quick single-point calibration for some parameters.
The calibration data are stored in the non volatile probe memory, so that the
same probe can be used with different meters without needing recalibration.
• To perform a calibration, select “Calibration”
in the main menu with the arrow keys and then
press OK.
NO
NO
NO
NO
NOTE
TE
TE
TE
TE
If the password is enabled and the latest
function displayed was not a password
protected feature, the meter will ask to
insert the password.
• Select the calibration type with the arrow keys and then press OK.
The available options are:
Quick calibration (single point procedure to calibrate the D.O. saturation, pH
and conductivity ranges), pH, D.O., conductivity, atmospheric pressure, ORP and
temperature.
4.1 QUICK CALIBRATION
The quick calibration feature allows a fast and easy calibration of the multipa-
rameter probe on the field, using only one solution (HI9828-25
HI9828-25
HI9828-25
HI9828-25
HI9828-25).
• Fill the calibration beaker with the HI9828-25
HI9828-25
HI9828-25
HI9828-25
HI9828-25 calibration solution.
• Screw the calibration beaker on the probe body. Some solution will overflow.
• Wait a few minutes to stabilize.
• In “Calibration” menu, select the “Quick cali-
bration” option and press OK.
• A 3-item (pH, Conductivity and Dissolved Oxy-
gen) screen appears, “pH” starts blinking and
the “Not ready” message is shown on the lower
part of the window.